IM
IronMonkey Threat Research

CVE-2026-7251 CRITICAL

Published: 2026-05-26 | Last Modified: 2026-05-26 | Status: Awaiting Analysis

Description

Eppendorf BioFlo 320 is vulnerable to due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.

CVSS Metrics

Base Score: 9.8 (CRITICAL)

CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

Attack VectorNETWORK
Attack ComplexityLOW
Privileges RequiredNONE
User InteractionNONE
ScopeUNCHANGED
Confidentiality ImpactHIGH
Integrity ImpactHIGH
Availability ImpactHIGH

Source: [email protected]

Type: Secondary

Exploitability Score: 3.9

Impact Score: 5.9

Base Score: 9.3 (CRITICAL)

CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X

Attack VectorNETWORK
Attack ComplexityLOW
Attack RequirementsNONE
Privileges RequiredNONE
User InteractionNONE
Vulnerability ConfidentialityHIGH
Vulnerability IntegrityHIGH
Vulnerability AvailabilityHIGH
Subsequent ConfidentialityNONE
Subsequent IntegrityNONE
Subsequent AvailabilityNONE

Source: [email protected]

Type: Secondary

Weaknesses

Source Type Description
[email protected] Primary
en CWE-259
Notification
Message here